Emergency Quick-Start
Press the master switch (1)
–
The pressure gauges
(2)
are in the green zone.
–
All lamps on the ventilator light up.
Press a function key on the ventilator
Recommendation: Press and hold down
M
key for
longer than 1 second.
–
The message »
Test interrupted
« is displayed.
For devices without PCV mode (software 6.n):
Deliver fresh gas
–
With the delivery valve buttons
(3)
under the
measuring tubes
–
If necessary fill the system rapidly with the O
2
flush
(4)
.
–
The manual ventilation bag should fill.
Set the APL pressure limiting valve (5) to »MAN«
Set maximum pressure
–
Turn the valve head
(6)
until the disc is at the
desired maximum pressure.