Uninterruptible Power Supply
(optional, see
Instructions for Use of the UPS)
When the machine is powered by the UPS battery
during a mains power failure, no power is supplied to
the auxiliary sockets!
The Desflurane Vapor unit is powered by the side
connector for non-heating apparatus (No. 9 on page 140)
and will therefore continue to run as normal.
1
Plug the Cicero power plug into the socket of the
UPS.
2
Plug the UPS mains plug into the mains socket.
The UPS can supply the Cicero EM with electrical energy
for about 45 minutes. It is activated automatically in the
event of a power failure.
In the case of a ceiling unit, the auxiliary sockets are
not fitted.
Connecting the compressed gas supply
3
Screw hoses for O
2
, AIR and N
2
O into the rear of the
apparatus and plug connectors into the wall sockets.
Switch-on lock:
If the machine is switched on without pressure on the
O
2
line, it is impossible to meter the other gases for
safety reasons!
»Pressure supply«
appears on the ventilator.
●
Connect O
2
supply and confirm.
●
Check that the supply pressure is adequate on the
pressure gauges on the front (pointers must be in the
green area).
4
Holder for anaesthetic gas scavenging. See next page
and separate Instructions for Use.
●
If the machine is equipped with a vacuum bronchial
aspirator (option), a vacuum supply system must be
connected.
Preparing for use
UPS
Compressed gas supply