Message, cause and remedy
Messages on the system screen
(In alphabetical order according to priority)
Messages
Cause
Remedy
PRESSURE INOP !
Pressure sensor defective.
Call DrägerService
RS 232 COM !
RS 232 communication interrupted.
Check plug connection.
SELECT AGAS !
The automatically recognised anaesthetic agent
does not match the manually selected anaesthetic.
Check the anaesthetic agent
used or switch over to automatic
recognition.
SPO2
G
OFF !
SpO
2
alarm inactive.
SPO2 INOP !
SpO
2
measurement faulty.
Call DrägerService
SPO2 SENSOR? !
No pulse detected by pulsoxymeter although heart
is clearly beating.
Check patient's condition
(disturbed circulation?)
Check SpO
2
sensor positioned
correctly.
TEMP1 CALIB !
TEMP2 CALIB !
Fault found during automatic internal accuracy test.
Call DrägerService
TEMP 1 INOP !
TEMP 2 INOP !
Temperature measuring function defective.
Replace temperature sensor.
Call DrägerService
WATER TRAP ? !
The water trap is full or severely fouled.
No measured values are displayed for CO
2
, N
2
O,
anaesthetic agent and O
2
.
Empty the water trap.
Check the fitting of the
collecting jar. Remove
condensate from the prism.