Fit the piston pump
●
Swing up table top
1
Handle in starting position -
Do not damage gaskets when fitting pump!
●
Hold piston pump by its handle and insert in machine.
2
Swing handle as far as possible to the left (black).
Fit the breathing system
3
Hold breathing system by its handle and insert in
machine -
4
Lock in direction of arrow -
swing as far as possible (black).
Do not damage gaskets when fitting breathing
system!
Fit valves
5
Fit valve discs in expiration and inspiration valves -
fit sealing rings - screw on valve caps.
6
Screw on pressure limitation valve APL -
scale facing the front.
Assemble and fit the inspiratory O
2
sensor
(if used)
●
Do not touch wire mesh on cell!
Insert cell with conductors facing wiring -
●
Screw sensor housing together -
7
Plug sensor onto inspiration valve.
Replace O
2
sensor for gas sampling system
(Only if measuring cell is exhausted; display
»
FI O
2
. INOP !
«)
8
Unscrew the thumbscrew on the right-hand panel of
the device by turning counter-clockwise -
●
Draw out old sensor and insert new sensor in mount -
●
Screw thumb screw back into place. The sensor is
automatically calibrated.
127
Reassembling machine
Reassembling the parts
2
1
6
4
3
5
G
u
E
S
Q
s
7
8
18029278
03329278
18129278
17329278
18229278