Check Vapor:
– Set the handwheel to »
O
«.
– Filling level OK?
– Last inspection less than six months ago?
Safety filler:
– The plug is inserted and secured.
Vaporiser mount:
– Vaporiser is correctly seated.
Locking mechanism:
– Plug-in system is locked (lever to the left as far as
possible).
Check anaesthetic gas scavenging:
– Is indicator in wall socket green? (Only when using
Dräger systems; note sounds of gas flow in other
cases.)
– Hose connected to breathing system?
– Is the anaesthetic gas scavenging system correctly
installed?
– Are the two hoses fitted correctly?
– Are the vents open?
– Are all hoses free of kinks?
1
Is the transfer hose of the anaesthetic gas scavenging
system free of condensate? Drain if necessary.
22
0
Vol. %
Preparing for use
Checking workstation with checklist
Check-list Cicero EM
Vapor handwheel set to zero
fill - level ok
less than 6 months since last inspection
correctly seated and locked
Safety-fill plug secured in place
To be verified by the user prior to each use
Safety f p ug secu ed n p ace
Gas scavenger operational
Breathing system complete, locked in place
Soda lime no colour change
Emergency breathing bag present, functional