Message, cause and remedy
Messages on the system screen
(In alphabetical order according to priority)
Messages
Cause
Remedy
Caution messages
AMV
<
!!
Minute volume below lower alarm limit.
Tube blocked / kinked?
Leak in breathing system?
Pressure limited ventilation?
Check tube and hoses.
Seal breathing system.
Adjust ventilation pattern.
AMV
>
!!
Minute volume exceeds upper alarm limit.
Correct tidal volume or
respiration rate on ventilator.
DIA NIBP
<
!!
Measured non-invasive blood pressure below
lower alarm limit.
Check patient's condition!
Correct alarm limit if necessary.
DIA NIBP
>
!!
Measured non-invasive blood pressure exceeds
upper alarm limit.
Check patient's condition!
Correct alarm limit if necessary.
DIAS IBP1
<
!!
DIAS IBP2
<
!!
Measured invasive diastolic pressure of the
channel shown (1 or 2) below lower alarm limit.
Check patient's condition!
Correct alarm limit if necessary.
DIAS IBP1
>
!!
DIAS IBP2
>
!!
Measured invasive diastolic pressure of the
channel shown (1 or 2) exceeds upper alarm limit.
Check patient's condition!
Correct alarm limit if necessary.
ET CO2
>
!!
Upper alarm limit for end-expiratory CO
2
concentration has been exceeded for at least two
breaths.
Check ventilation
ET CO2
<
!!
End-expiratory CO
2
concentration below lower
alarm limit for at least two breaths.
Check ventilation.
Check for leak in sample line.
EXSP-V INOP !!
Fault in expiration valve. Measured value for minute
ventilation may be excessive.
Check expiration valve.
Carry out another leak test.
Fl O2
>
!!
Inspiratory O
2
concentration exceeds upper alarm
limit.
Check O
2
concentration in fresh
gas flow.
This message is deactivated
until the next leakage test by
pressing the alarm suppressor
key.
Fl HAL
<
!!
Fl ISO
<
!!
Fl ENF
<
!!
Fl DES
<
!!
Fl SEV
<
!!
Applicable inspiratory anaesthetic concentration
below lower alarm limit.
Value below lower alarm limit for at least two
breaths.
Check setting of anaesthetic
vaporizer.
FRESH GAS ? !!
Too little gas in manual ventilation bag.
Increase fresh gas flow. Repair
any leaks in hose system.
HEART RATE
>
!!
Heart rate exceeds set upper limit.
Check patient's condition!
Correct alarm limit if necessary.
INSP CO2
>
!!
The inspiratory CO
2
concentration exceeds the
upper alarm limit.
Check the soda lime.
Check the valve plate in the
breathing system.
Correct alarm limit if necessary.