Message, cause and remedy
Messages in the ventilation system
Message must be acknowledged
Messages during
the self-test
Cause
Remedy
Service No. 118
• Incorrect test control pressure. No sealing
Call DrägerService.
test possible for valve and roller diaphragm.
Service No. 119
• Test control pressure too high.
Call DrägerService.
Service No. 120
• Operating control pressure too high.
Call DrägerService.
Service No. 214
• Cold/warm restart recognition not
Call DrägerService.
guaranteed.
Standby
The machine is ready for operation.
Select operating mode.
Subsystem 1/2 leak
• Leak in subsystem 1 and/or 2.
Check subsystems.
(Leak greater than 5 L/min at 30 mbar)
(see chart on page 164).
Subsystem 1 leak
• Leak in area of inspiration valve, valve 7,
Close the quick-release couplings.
patient hose, flow sensor or O
2
sensor
Connect the P-sensor. Replace the valve disc
and PEEP valve (V
6
)
in the inspiration valve.
Subsystem 2 leak
• Leak in area of inspiration valve, absorber,
Close quick-release couplings.
piston pump, valves 1-2-3, piston
pump/breathing system interface and
PEEP valve (V
6
)
Subsystem 3 leak
• Leak in area of breathing bag,
Check connections and breathing
fresh gas hose, Vapor
bag or hoses. Check connector system and
sealing rings on Vapor.
Test ..x interrupted
Displayed after the self-test is stopped.
Select operating mode.
Valve plate faulty
• Breathing system valve plate or
Replace breathing system or valve plate.
diaphragm of V
1
defective.
Pneumatic interface faulty (check lip seals)
Display
»112«
in V
T
field
Valve plate faulty
• Breathing system valve plate or
Replace breathing system or valve plate.
diaphragm of V
2
defective.
Pneumatic interface faulty (check lip seals).
Display
»113«
in V
T
field
Valve plate faulty
• Breathing system valve plate or
Replace breathing system or valve plate.
diaphragm of V
3
defective.
Pneumatic interface faulty (check lip seals).
Display
»114«
in V
T
field
Version ....
Display at the start of the self-test
Software version No. is displayed after
switching on. In the event of malfunctions,
quote the version number to DrägerService.
Y-piece closed ?
• Dialogue during the self-test
Close the Y-piece for the remainder of the
self-test.
Y-piece open ?
• Dialogue during the self-test
Open the Y-piece for the internal pressure
sensor calibration. If using an external
humidifier, remove the breathing hose from
the inspiration cone or filter.
????
Internal electronic malfunction.
Call DrägerService.
Last interruption
The self-test has already been stopped
For information. The next time you start up the
9 times!
machine, the complete self-test will have to
be run.
113