Emergency Quick-Start
7
Notes on Quick Start
The Quick Start procedure is permitted a maximum of
10 times in succession, and only if the previous (full) self-
test was completed without error.
Otherwise, the display on the ventilator and the system
screen instruct the user to run through the complete self-
test.
Manual ventilation remains possible under all condi-
tions!
●
Quick Start can also be triggered by the
»Cancel
Test«
softkey.
●
Quick Start can be triggered at any time, even during
a self-test.
If there is a power failure
(manual ventilation is
possible)
Make sure that the master switch is pressed down
–
The acoustic power failure warning is muted after
30 seconds.
Deliver fresh gas - set APL valve
–
If required, press the O
2
flush button (»
O
2
+
«).
Note: If there is a power failure, the piston of the
ventilator is pushed to its end position by the
airway pressure. This increases the system volume
by a maximum of 1.4 litres.
If there is a gas failure
In case of AIR failure
(medical air)
–
The system automatically switches over to O
2
supply.
In case of O
2
failure
–
The system automatically switches over to the
AIR supply. An acoustic alarm is sounded (O
2
shortage signal). The supply of N
2
O is disabled.
If both O
2
and air fail
The patient must immediately be ventilated with the
separate emergency ventilation bag!
Self-test
16:47
00:00
cancel
test
Self-test cancelled
06-12-98
00429278