Transport function
The parameter box stores the following data, which can
be transferred to the system screen:
– Patient data from the list screen,
– Alarm limit settings and
– Measurement parameter settings.
The patient remains connected to the parameter box
while in transit. The patient's data remains constantly
available. There is no need to reconnect the patient and
set system screens.
The transport function can be preset in
»Standby«
(operation in normal use is described on page 83 in the
»Parameter box«
section).
Transport function on/off
When switched
»off«
, no data is exchanged between
the parameter box and the system screen (suitable for
stationary use of the parameter box).
When switched
»on«
, data is exchanged.
Transport menus yes/no
If switched to
»no«
, data is transferred between the
parameter box and the system screen in accordance
with the presetting in
»Standby«
.
If switched to
»yes«
, the user can modify the type and
scope of data transfer during operation (see also
»Parameter box«
section on page 86).
The patient is connected for the first time to …
If a patient is connected for the first time, the parameter
box will not yet contain any stored data for the patient.
This will generally be the case in the admissions room.
If so, enter
»yes«
next to
»parameter box«
.
In response, any remaining data from the previous
patient will be deleted!
If the patient arrives with parameter box already
connected, i.e. with a "previous history" on the device,
enter
»no«
after
»Parameter box«
. Data will then be
transferred from the parameter box to the system screen.
Such cases will generally occur in the operating theatre
and in the recovery room.
62
System screen functions
Configuring the system screen in »Standby«
Select a menu position. Please confirm !
Change the settings. Please confirm again !
Standby / Configuration
11:59
00:00
alarms inactive !
Settings
mode
adult neon.
alarm limits
parameter
screen
acoustic
list entry
transport
RS 232 (Medibus)
record (printer)
basic config.
transport
function on off
transport-
menus yes no
The patient is new
to the:
monitor yes no
parameter
box yes no
Which data do you want
to transfer from the
parameter box?
settings yes no
alarm limits yes no
list yes no
06-12-98
11829278