27
LEEP System 1000
®
Workstation • Model KH1000A • Operating Manual • English
(continued)
Recommended separation distance between portable and mobile RF communications equipment
and the LEEP System 1000
®
Electrosurgical Generator.
The Electrosurgical Generator is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Electrosurgical Generator can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the Electrosurgical Generator as recommended below, according to the maximum
output power of the communications equipment.
Section 19 Specifications
19.1 Electrical – LEEP System 1000
®
Workstation
INPUT VOLTAGE . . . . . . . . . . . . . . . . . . 190 –250 VAC, 50 Hz
CURRENT . . . . . . . . . . . . . . . . . . . . . . . 3.0 amps max.
FUSES . . . . . . . . . . . . . . . . . . . . . . . . . . 8.0 amps, 250 V, T type
WORKING FREQUENCY . . . . . . . . . . . . 450 kHz
OUTPUT POWER . . . . . . . . . . . . . . . . . 100 watts RMS (500 ohm load)
DUTY CYCLE . . . . . . . . . . . . . . . . . . . . . Intermittent 10/30 seconds
LOW FREQUENCY LEAKAGE . . . . . . . Less than 50 micro-amps
POWER CORD . . . . . . . . . . . . . . . . . . . . Various
WEIGHT . . . . . . . . . . . . . . . . . . . . . . . . . Approximately 98 pounds (44 kg)
Rated maximum
output power
of transmitter
Watts
Separation distance according
to frequency of transmitter
Meters
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.1167
0.1167
0.2333
0.1
0.3689
0.3689
0.7379
1
1.1667
1.1667
2.3333
10
3.6894
3.6894
7.3789
100
11.667
11.667
23.333
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output rating of the transmitter in watts (W) according to the
transmitter manufacturer.
For the Electrosurgical Generator,
V
1
= 3 Vrms
E
1
= 3 V/m
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Содержание LEEP System 1000
Страница 3: ...37864 Rev A 4 13 LEEP System 1000 Workstation 220 VAC Model KH1000A 1 Operating Manual English ...
Страница 4: ...This page left blank intentionally 2 ...
Страница 34: ...NOTES 32 LEEP System 1000 Workstation Model KH1000A Operating Manual English continued ...
Страница 35: ...37864 Rev A 4 13 Workstation des LEEP System 1000 220 V AC Modell KH1000A 33 Bedienungshandbuch Deutsch ...
Страница 36: ...Leerseite 34 ...
Страница 66: ...NOTIZEN 64 Workstation des LEEP System 1000 Modell KH1000A Bedienungsanleitung Deutsch German Forts ...
Страница 68: ...Esta página se ha dejado en blanco intencionadamente 66 ...
Страница 99: ...37864 Rév A 4 13 Poste de travail pour système LEEP 1000 220 V Modèle KH1000A 97 Manuel d utilisation Français ...
Страница 100: ...Page intentionnellement vierge 98 ...
Страница 130: ...REMARQUES 128 Poste de travail pour système LEEP 1000 Modèle KH1000A Manuel d utilisation Français French suite ...
Страница 131: ...37864 Rev A 4 13 Stazione di lavoro LEEP System 1000 220 V c a Modello KH1000A 129 Manuale di funzionamento Italiano ...
Страница 132: ...Questa pagina è stata lasciata in bianco intenzionalmente 130 ...
Страница 162: ...NOTE 160 Stazione di lavoro LEEP System 1000 Modello KH1000A Manuale di funzionamento Italiano Italian continuazione ...
Страница 163: ...37864 Rev A 4 13 Estação de Trabalho LEEP System 1000 220 V CA Modelo KH1000A 161 Manual de Funcionamento Português ...
Страница 164: ...Página intencionalmente em branco 162 ...
Страница 194: ...NOTAS 192 Estação de Trabalho LEEP System 1000 Modelo KH1000A Manual de funcionamento Português Portuguese continuação ...
Страница 195: ......