17
LEEP System 1000
®
Workstation • Model KH1000A • Operating Manual • English
(continued)
By varying the pulse to an intermediate duration, it is possible to get a clinical effect that combines,
or “blends”, the clinical characteristics of CUT and COAG, yielding the effect referred to as
“BLEND”, where tissue is divided with a desirable amount of hemostatis along the margins of the
divided tissue.
The electrosurgical effect may vary throughout the procedure, requiring the operator to adjust the
relative power setting of the Electrosurgical Generator.
6.1.3 Select the Output Mode (i.e., “CUT”, “BLEND” or “COAG”, by Pushing the Corresponding Buttons
Output Mode Waveform Description
General Effect
CUT Continuous 450 kHz sinusoid
Cutting without Hemostatis
with minimal modulation
BLEND Interrupted 450 kHz sinusoid
Cutting with minimal
intermediate-duty cycle
Hemostatis
COAG Bursts of 450 kHz sinusoid
Coagulation without Cutting
short-duty cycle
6.1.4 Set the level of Output Power (confirmed on the digital display) by using the Output Power
Selector Buttons as Desired
WARNING
The degree and speed of electrosurgical effect is largely dependent on Current Density
at the point of contact of the active electrode. Loop electrosurgical excision procedure
electrodes from other manufacturers may vary in the diameter thickness, size and
configuration of the cutting wire. This may result in SIGNIFICANT changes in the
electrosurgical effect at a given output power level setting. The use of CooperSurgical
LEEP Electrodes is recommended.
6.2 Guidelines for Power Settings
The following guidelines for power settings may vary due to the technique, clinical circumstances,
accessory style, cutting wire diameter, size, configuration and user preference.
Recommended Power Settings (Watts) for the CooperSurgical LEEP System 1000
®
Electrodes
Wattage Suggestions for Fischer Cone Biopsy Excisor™
The above settings are for reference only and can be varied based on specific situations and the
experience of the operator.
NOTE
*
If the cut mode is desired,
use the recommended
settings for the blend mode.
+
The power setting can be
increased beyond 56 to
coagulate any bleeding points
as required.
Style
Loop Width (cm)
Ball
Electrodes
Needle
Electrodes
1.0
1.5
2.0
Macro
Micro
CUT
*
—
—
—
—
—
14–24
BLEND
22–36
30–40
34–46
—
—
—
COAG
—
—
—
40–56
+
30–36
—
Part
Number
Size
Wattage Rating
Part
Number
Size
Wattage Rating
900-150
Small
30-35 Watts
900-152
Large
45-50 Watts
900-151
Medium
40-45 Watts
900-157
Small Wide
40-45 Watts
900-154
Medium
Extended
40-45 Watts
900-158
Small Wide
Extended
40-45 Watts
900-155
Large
Shallow
40-45 Watts
Содержание LEEP System 1000
Страница 3: ...37864 Rev A 4 13 LEEP System 1000 Workstation 220 VAC Model KH1000A 1 Operating Manual English ...
Страница 4: ...This page left blank intentionally 2 ...
Страница 34: ...NOTES 32 LEEP System 1000 Workstation Model KH1000A Operating Manual English continued ...
Страница 35: ...37864 Rev A 4 13 Workstation des LEEP System 1000 220 V AC Modell KH1000A 33 Bedienungshandbuch Deutsch ...
Страница 36: ...Leerseite 34 ...
Страница 66: ...NOTIZEN 64 Workstation des LEEP System 1000 Modell KH1000A Bedienungsanleitung Deutsch German Forts ...
Страница 68: ...Esta página se ha dejado en blanco intencionadamente 66 ...
Страница 99: ...37864 Rév A 4 13 Poste de travail pour système LEEP 1000 220 V Modèle KH1000A 97 Manuel d utilisation Français ...
Страница 100: ...Page intentionnellement vierge 98 ...
Страница 130: ...REMARQUES 128 Poste de travail pour système LEEP 1000 Modèle KH1000A Manuel d utilisation Français French suite ...
Страница 131: ...37864 Rev A 4 13 Stazione di lavoro LEEP System 1000 220 V c a Modello KH1000A 129 Manuale di funzionamento Italiano ...
Страница 132: ...Questa pagina è stata lasciata in bianco intenzionalmente 130 ...
Страница 162: ...NOTE 160 Stazione di lavoro LEEP System 1000 Modello KH1000A Manuale di funzionamento Italiano Italian continuazione ...
Страница 163: ...37864 Rev A 4 13 Estação de Trabalho LEEP System 1000 220 V CA Modelo KH1000A 161 Manual de Funcionamento Português ...
Страница 164: ...Página intencionalmente em branco 162 ...
Страница 194: ...NOTAS 192 Estação de Trabalho LEEP System 1000 Modelo KH1000A Manual de funcionamento Português Portuguese continuação ...
Страница 195: ......