Section 11 Practical Suggestions
To optimize the performance when using an electrosurgical unit, the active electrode must be kept clean and
use the lowest possible power setting required. Some sparks or superficial carbonization of the tissue may
occur and the delivered power may decrease as a result of the electrical insulation caused by the tissue
charring.
Too high of a power setting results in a shorter surgical procedure, but may cause discharges and/or
superficial carbonization, sparking, arcs, etc.
Section 12 Cleaning the LEEP System 1000
®
Electrosurgical
Generator and Smoke Evacuator
LEEP System 1000 Electrosurgical Generator:
The Electrosurgical Generator may be cleaned with mild
soap solution – but be sure that fluid does not enter the system. Wipe dry.
Smoke Evacuator:
The outside of the Smoke Evacuator may be cleaned as necessary with a soft cloth
dampened (not wet) with isopropyl alcohol. Just as for the Electrosurgical Generator, do not allow liquid
to get inside of the unit.
Section 13 Troubleshooting
In the event of a failure during SELF-TEST, the display will show one of the error codes listed below:
ALARM SIGNALS
PROBLEM AUDIBLE TONE DISPLAY
RAM Memory 1 kHz 100ms ON 250ms OFF “Er0”
(during self-test phase)
EPROM Memory 1 kHz 100ms ON 250ms OFF “Er1”
(during self-test phase)
Signal Modulation 1 kHz 200ms ON 250ms OFF “Er2”
(during self-test phase)
Signal Modulation 1 kHz 40ms ON 60ms OFF “Er2”
(during activation phase)
Supply Voltage 1 kHz 100ms ON 250ms OFF “Er3”
(during self-test phase)
Supply Voltage 1 kHz 40ms ON 60ms OFF “Er3”
(during activation phase)
Power Output 1 kHz 100ms ON 250ms OFF “Er5”
(during activation phase)
Foot Pedal Circuit 1 kHz 100ms ON 250ms OFF “Er6”
(during self-test phase)
Microcontroller Power Supply 1 kHz 100ms ON 250ms OFF “Er7”
Dispersive Electrode 1 kHz 80ms ON 125ms OFF “nP”
& Red Light
A. If upon completion of Self-Test the Electrosurgical Generator display reads Er0, Er1, Er2, Er3, Er5, Er6
or Er7, it must be returned to CooperSurgical for repair (refer to Section 15).
B. If upon completion of Self-Test or before, if the Electrosurgical Generator stops functioning and the
display reads nP and the red warning light is on, check the Dispersive Patient Plate to ensure that it
is properly connected to the Electrosurgical Generator.
C. If, following correct set-up of the system, it does not operate, works in an intermittent way, or stops
working after a few seconds (without an audible signal), check for correct connection of pedal and its
condition. The pedal is pneumatic, so even a slight leak can cause a performance problem.
22
LEEP System 1000
®
Workstation • Model KH1000A • Operating Manual • English
(continued)
Содержание LEEP System 1000
Страница 3: ...37864 Rev A 4 13 LEEP System 1000 Workstation 220 VAC Model KH1000A 1 Operating Manual English ...
Страница 4: ...This page left blank intentionally 2 ...
Страница 34: ...NOTES 32 LEEP System 1000 Workstation Model KH1000A Operating Manual English continued ...
Страница 35: ...37864 Rev A 4 13 Workstation des LEEP System 1000 220 V AC Modell KH1000A 33 Bedienungshandbuch Deutsch ...
Страница 36: ...Leerseite 34 ...
Страница 66: ...NOTIZEN 64 Workstation des LEEP System 1000 Modell KH1000A Bedienungsanleitung Deutsch German Forts ...
Страница 68: ...Esta página se ha dejado en blanco intencionadamente 66 ...
Страница 99: ...37864 Rév A 4 13 Poste de travail pour système LEEP 1000 220 V Modèle KH1000A 97 Manuel d utilisation Français ...
Страница 100: ...Page intentionnellement vierge 98 ...
Страница 130: ...REMARQUES 128 Poste de travail pour système LEEP 1000 Modèle KH1000A Manuel d utilisation Français French suite ...
Страница 131: ...37864 Rev A 4 13 Stazione di lavoro LEEP System 1000 220 V c a Modello KH1000A 129 Manuale di funzionamento Italiano ...
Страница 132: ...Questa pagina è stata lasciata in bianco intenzionalmente 130 ...
Страница 162: ...NOTE 160 Stazione di lavoro LEEP System 1000 Modello KH1000A Manuale di funzionamento Italiano Italian continuazione ...
Страница 163: ...37864 Rev A 4 13 Estação de Trabalho LEEP System 1000 220 V CA Modelo KH1000A 161 Manual de Funcionamento Português ...
Страница 164: ...Página intencionalmente em branco 162 ...
Страница 194: ...NOTAS 192 Estação de Trabalho LEEP System 1000 Modelo KH1000A Manual de funcionamento Português Portuguese continuação ...
Страница 195: ......