15
LEEP System 1000
®
Workstation • Model KH1000A • Operating Manual • English
(continued)
6.
When this equipment is operated:
a.
A Patient Return Pad (dispersive pad) of adequate surface area MUST be properly attached to
the patient or the risk of accidental burns will exist.
b.
The Patient Return Pad (dispersive pad) should be placed as close as possible to the site of
use of the active accessory – but MUST NEVER be placed so as to allow the patient’s heart to
be in the pathway from the active accessory to the return electrode!
7.
The user should thoroughly understand the principles and use of radio frequency (RF) current before
using this equipment. This understanding is essential to avoid the hazard of shocks or burns to the
user and/or the patient.
8.
The instructions for use described in this manual must be followed; otherwise, compromised safety,
malfunction, injury to the operator and/or patient, or costly damage to the Electrosurgical Generator
may occur.
9.
There are no user-serviceable parts within the housing. Repairs to this equipment should only be
performed by authorized CooperSurgical service personnel. For service information, please contact
CooperSurgical (refer to Section 15).
5.6 Electrosurgical Procedures
This section provides only general information about the use of electrosurgical devices. Only the user
can evaluate the clinical factors involved with each patient and determine if the use of this equipment is
indicated. The user must then decide on the specific technique and procedure that will accomplish the
desired clinical effect.
WARNING
Electrosurgical generators are designed to allow the controlled destruction of tissue and are
inherently dangerous if operated improperly.
REPORTED PROBLEMS DUE TO IMPROPER OPERATION DURING ELECTROSURGICAL
PROCEDURES HAVE INCLUDED:
•
Inadvertent activation with resultant tissue damage at the wrong site and/or
equipment damage
•
Alternate current pathways resulting in burns where the patient or physician or assistant
is in contact with exposed metal
•
Explosions caused by electrosurgical sparking in a flammable gas mixture (i.e., explosive
anesthetic gases and the inappropriate use of alcohol and other flammable liquids)
•
Perforation and massive hemorrhage
A proper patient return pad pathway is extremely important during any monopolar
electrosurgical procedure. Every effort must be made to ensure that, throughout the
electrosurgical procedure, an adequate surface area is provided and remains in proper
contact with the patient to reduce the current density below a level that might cause
inadvertent tissue damage where the patient return pad has been applied.
5.7 Turning on the Smoke Evacuator and LEEP System 1000
®
Electrosurgical Generator
NOTE:
Numbers in
blue boxes
refer to those component parts shown in Figure 6.
Push the POWER switch
, located under the Electrosurgical Generator (for the Smoke Evacuator), to
the “On” position. This will start the Smoke Evacuator.
Turn ON the Electrosurgical Generator using the power switch
. The Electrosurgical Generator
automatically performs a SELF-TEST that checks RAM memory, EPROM memory, supply voltage, signal
modulation and the following displays: function selector green lights, digital display, cut and coagulant yellow
light, coagulation blue light, and the audible signal. When the Electrosurgical Generator passes SELF-TEST,
the display will show current software revision, i.e., r2A, r2B, for several seconds, then will go blank.
1
12
Содержание LEEP System 1000
Страница 3: ...37864 Rev A 4 13 LEEP System 1000 Workstation 220 VAC Model KH1000A 1 Operating Manual English ...
Страница 4: ...This page left blank intentionally 2 ...
Страница 34: ...NOTES 32 LEEP System 1000 Workstation Model KH1000A Operating Manual English continued ...
Страница 35: ...37864 Rev A 4 13 Workstation des LEEP System 1000 220 V AC Modell KH1000A 33 Bedienungshandbuch Deutsch ...
Страница 36: ...Leerseite 34 ...
Страница 66: ...NOTIZEN 64 Workstation des LEEP System 1000 Modell KH1000A Bedienungsanleitung Deutsch German Forts ...
Страница 68: ...Esta página se ha dejado en blanco intencionadamente 66 ...
Страница 99: ...37864 Rév A 4 13 Poste de travail pour système LEEP 1000 220 V Modèle KH1000A 97 Manuel d utilisation Français ...
Страница 100: ...Page intentionnellement vierge 98 ...
Страница 130: ...REMARQUES 128 Poste de travail pour système LEEP 1000 Modèle KH1000A Manuel d utilisation Français French suite ...
Страница 131: ...37864 Rev A 4 13 Stazione di lavoro LEEP System 1000 220 V c a Modello KH1000A 129 Manuale di funzionamento Italiano ...
Страница 132: ...Questa pagina è stata lasciata in bianco intenzionalmente 130 ...
Страница 162: ...NOTE 160 Stazione di lavoro LEEP System 1000 Modello KH1000A Manuale di funzionamento Italiano Italian continuazione ...
Страница 163: ...37864 Rev A 4 13 Estação de Trabalho LEEP System 1000 220 V CA Modelo KH1000A 161 Manual de Funcionamento Português ...
Страница 164: ...Página intencionalmente em branco 162 ...
Страница 194: ...NOTAS 192 Estação de Trabalho LEEP System 1000 Modelo KH1000A Manual de funcionamento Português Portuguese continuação ...
Страница 195: ......