Section 8. Operation
420
Note C terminals configured as Tx transmit only 0 to 5 Vdc logic.
However, C terminals configured as Rx read most true RS-232 signals.
When connecting serial sensors to a C terminal configured as Rx, the
sensor power consumption may increase by a few milliamps due to
voltage clamps in the CR3000. An external resistor may need to be added
in series to the Rx line to limit the current drain, although this is not
advisable at very high baud rates. See Circuit to Limit C Terminal Input to
5 Volts
(p. 420).
FIGURE 101: Circuit to Limit C Terminal Input to 5 Vdc
8.1.6.2 SDI-12 Sensor Support — Details
Related Topics:
• SDI-12 Sensor Support — Overview
(p. 78)
• SDI-12 Sensor Support — Details
(p. 420)
• Serial I/O: SDI-12 Sensor Support — Programming Resource
(p. 255)
SDI-12 is a communication protocol developed to transmit digital data from smart
sensors to data-acquisition units. It is a simple protocol, requiring only a single
communication wire. Typically, the data-acquisition unit also supplies power (12
Vdc and ground) to the SDI-12 sensor. SDI12Recorder() instruction
communicates with SDI-12 sensors on terminals configured for SDI-12 input.
See the table CR3000 Terminal Definitions
to determine those terminals
configurable for SDI-12 communications.
8.1.7 Field Calibration — Overview
Related Topics:
• Field Calibration — Overview
(p. 79)
• Field Calibration — Details
(p. 229)
Calibration increases accuracy of a measurement device by adjusting its output, or
the measurement of its output, to match independently verified quantities.
Adjusting sensor output directly is preferred, but not always possible or practical.
By adding FieldCal() or FieldCalStrain() instructions to the CR3000 CRBasic
program, measurements of a linear sensor can be adjusted by modifying the
programmed multiplier and offset applied to the measurement without modifying
or recompiling the CRBasic program.
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......