Appendix A. Info Tables and Settings
580
CSIOInfo
String
•
Settings Editor: CS I/O | {info box}
2
Order and definitions of auto self-calibration array elements:
(1) 5000 mV range 250 ms integration
(2) 1000 mV range 250 ms integration
(3) 200 mV range 250 ms integration
(4) 50 mV range 250 ms integration
(5) 20 mV range 250 ms integration
(6) 5000 mV range 60 Hz integration
(7) 1000 mV range 60 Hz integration
(8) 200 mV range 60 Hz integration
(9) 50 mV range 60 Hz integration
(10) 20 mV range 60 Hz integration
(11) 5000 mV range 50 Hz integration
(12) 1000 mV range 50 Hz integration
(13) 200 mV range 50 Hz integration
(14) 50 mV range 50 Hz integration
(15) 20 mV range 50 Hz integration
Info Tables and Settings: D
Keyword
Data Type
R
ead
O
n
ly
• Where to Find
Description
DataFillDays()
Numeric
Y
•
DataTableInfo table
Reports the time required to fill a data table. Each table has its own entry in a two-
dimensional array. First dimension is for on-board memory. Second dimension is for card
memory.
DataRecord
Size()
Numeric
Y
•
DataTableInfo table
Reports the number of records in a data table.
DataTable
Name()
String
Y
•
DataTableInfo table
Reports the names of data tables. Array elements are in the order the data tables are
declared in the CRBasic program.
DeleteCardFiles
OnMismatch
Numeric
•
Keyboard: Settings (Advanced)
Defaults to False
DNS()
String
•
Settings Editor: Ethernet, CS I/O IP, PPP, Wi-Fi | DNS Server 1, DNS Server 2
•
Keyboard: Settings (Network Services)
Specifies the addresses of two domain name servers that the CR3000 can use to resolve
domain names to IP addresses. Note that if DHCP is used to resolve IP information, the
addresses obtained via DHCP are appended to this list. Defaults to 0.0.0.0. Form: 0–255.0–
255.0–255.0-255
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......