Section 8. Operation
362
Input Error Examples: Type T Thermocouple @ 45°C
These examples demonstrate that in the environmental temperature range, input-
offset error is much greater than input-gain error because a small input range is
used.
Conditions:
CR3000 module temperature, –25 to 50 °C
Temperature = 45 °C
Reference temperature = 25 °C
Delta T (temperature difference) = 20 °C
Thermocouple output multiplier at 45 °C = 42.4 µV °C
-1
Thermocouple output = 20°C • 42.4 µV °C
-1
= 830.7 µV
Input range = ±20 mV
Error Calculations with Input Reversal = True
µV error = gain term + offset term
= (830.7 µV • 0.07%) + (1.5 • 0.67 µV + 1.0 µV)
= 0.581 µV + 2.01 µV
= 2.59 µV (= 0.061 °C)
Error Calculations with Input Reversal = False
µV Error = gain term + offset term
= (830.7 µV • 0.07%) + (3
• 0.67 µV + 2.0 µV)
= 0.581 µV + 4.01 µV
= 4.59 µV (= 0.11 °C)
Input Error Examples: Type K Thermocouple @ 1300°C
Error in the temperature due to inaccuracy in the measurement of the
thermocouple voltage increases at temperature extremes, particularly when the
temperature and thermocouple type require using the ±200|250 mV range. For
example, assume type K (chromel-alumel) thermocouples are used to measure
temperatures around 1300°C.
These examples demonstrate that at temperature extremes, input offset error is
much less than input gain error because the use of a larger input range is required.
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......