Section 8. Operation
402
Analog Voltage Measurement Offsets
Differential
Measurement
With Input Reversal
Differential
Measurement
Without Input
Reversal
Single-Ended
1.5 • Basic Reso
1.0
µ
V
3 • Basic Reso
2.0
µ
V
3 • Basic Reso
3.0
µ
V
Note — the value for Basic Resolution is found in the table Analog Voltage
Measurement Resolution
(p. 352).
Analog Voltage Measurement Resolution
Input
Voltage Range
(mV)
Differential
Measurement
With Input Reversal
(
µ
V)
Basic Resolution
(
µ
V)
±
5000
83.33
167
±
1000
16.67
33.4
±
200
3.33
6.67
±
50
0.83
1.67
±
20
0.33
0.67
Note — see Specifications
(p. 95)
for a complete tabulation of measurement
resolution
As an example, figure Voltage Measurement Accuracy Band Example
(p. 354)
shows changes in accuracy as input voltage changes on the ±1000 and ±5000
input ranges. Percent-of-reading is the principle component, so accuracy
improves as input voltage decreases. Offset is small, but could be significant in
applications wherein the sensor-signal voltage is very small, such as is
encountered with thermocouples.
Offset depends on measurement type and voltage-input range. Offsets equations
are tabulated in table Analog Voltage Measurement Offsets
For example,
for a differential measurement with input reversal on the ±5000 mV input range,
the offset voltage is calculated as follows:
offset
= 1.5 • Basic Reso 1.0 µV
= (1.5 • 83.33 µV) + 1.0 µV
= 126 µV
where Basic Resolution is taken from the table Analog Voltage Measurement
Resolution
(p. 352).
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......