Section 7. Installation
231
•
Valid mode variable entries are 1 or 4.
Before, during, and after calibration, one of the following codes will be stored in
the CalMode variable:
FieldCal() Codes
Value Returned
State
-1
Error in the calibration setup
-2
Multiplier set to 0 or NAN
; measurement =
NAN
-3
Reps is set to a value other than 1 or the size of
MeasureVar
0
No calibration
1
Ready to calculate (KnownVar holds the first of a
two point calibration)
2
Working
3
First point done (only applicable for two point
calibrations)
4
Ready to calculate (KnownVar holds the second of a
two-point calibration)
5
Working (only applicable for two point calibrations)
6
Calibration complete
7.7.12.4.1
One-Point Calibrations (Zero or Offset)
Zero operation applies an offset of equal magnitude but opposite sign. For
example, when performing a zeroing operation on a measurement of 15.3, the
value –15.3 will be added to subsequent measurements.
Offset operation applies an offset of equal magnitude and same sign. For
example, when performing an offset operation on a measurement of 15.3, the
value 15.3 will be added to subsequent measurements.
See FieldCal() Zero or Tare (Opt 0) Example
(p. 233)
and FieldCal() Offset (Opt 1)
Example
(p. 235)
for demonstration programs:
1. Use a separate FieldCal() instruction and variables for each sensor to be
calibrated. In the CRBasic program , put the FieldCal() instruction
immediately below the associated measurement instruction.
2.
Set mode variable = 0 or 6 before
starting.
3. Place the sensor into zeroing or offset condition.
4. Set KnownVar variable to the offset or zero value.
5.
Set mode variable =
1 to start calibration.
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......