7.7.12.4.2
Two-Point Calibrations (gain and offset)
Use this two-point calibration procedure to adjust multipliers (slopes) and offsets
(y intercepts). See FieldCal() Slope and Offset (Opt 2) Example
(p. 238)
and
FieldCal() Slope (Opt 3) Example
(p. 240)
for demonstration programs:
1. Use a separate FieldCal() instruction and separate variables for each sensor to
be calibrated.
2.
Ensure mode variable =
0 or 6 before starting.
a. If Mode > 0
and ≠
6, calibration is in progress.
b. If Mode < 0, calibration encountered an error.
3. Place sensor into first known point condition.
4. Set KnownVar variable to first known point.
5. Set Mode
variable =
1 to start first part of calibration.
a. Mode
=
2 (automatic) during the first point calibration.
b. Mode
=
3 (automatic) when the first point is completed.
6. Place sensor into second known point condition.
7. Set KnownVar variable to second known point.
8. Set Mode
=
4 to start second part of calibration.
a. Mode
=
5 (automatic) during second point calibration.
b. Mode
=
6 (automatic) when calibration is complete.
7.7.12.4.3
Zero Basis Point Calibration
Zero-basis calibration (FieldCal() instruction Option 4) is designed for use with
static vibrating wire measurements. It loads values into zero-point variables
to track conditions at the time of the zero calibration. See FieldCal() Zero
Basis (Opt 4) Example
(p. 243)
for a demonstration program.
7.7.12.5 Field Calibration Examples
FieldCal() has the following calibration options:
•
Zero
•
Offset
•
Two-point slope and offset
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......