Section 7. Installation
244
FieldCalStrain() with the manufacturer's gage factor (GF), becoming the
adjusted gage factor (GF
adj
), which is then used as the gage factor in
StrainCalc(). GF is stored in the CAL file and continues to be used in
subsequent calibrations. Non-linearity of the bridge is compensated for by
selecting a shunt resistor with a value that best simulates a measurement near
the range of measurements to be made. Strain-gage manufacturers typically
specify and supply a range of resistors available for shunt calibration.
3. Shunt calibration verifies the function of the CR3000.
4. The zero function of FieldCalStrain() allows a particular strain to be set as an
arbitrary zero, if desired. Zeroing is normally done after the shunt calibration.
Zero and shunt options can be combined ina single CRBasic program.
CRBasic example FieldCalStrain() Calibration
(p. 245)
is provided to demonstrate
use of FieldCalStrain() features. If a strain gage configured as shown in figure
Quarter-Bridge Strain Gage with RC Resistor Shunt
(p. 245)
is not available, strain
signals can be simulated by building the simple circuit, substituting a 1000 Ω
potentiometer for the strain gage. To reset calibration tests, use the support
software File Control
(p. 539)
menu to delete .cal files, and then send the
demonstration program again to the CR3000.
Example Case: A 1000 Ω strain gage is placed into a resistiv
e bridge at position
R1. The resulting circuit is a quarter-bridge strain gage with alternate shunt-
resistor (Rc) positions shown. Gage specifications indicate that the gage factor is
2.0 and that with a 249 kΩ shunt, measurement should be about 2000 micro
strain.
Send CRBasic example FieldCalStrain() Calibration
(p. 245)
as a program to a
CR3000 datalogger.
7.7.12.6.2
FieldCalStrain() Shunt Calibration Example
CRBasic example FieldCalStrain() Calibration
(p. 245)
is provided to demonstrate
use of FieldCalStrain() features. If a strain gage configured as shown in figure
Quarter-Bridge Strain Gage with RC Resistor Shunt
(p. 245)
is not available, strain
signals can be simulated by building the simple circuit, substituting a 1000 Ω
potentiometer for the strain gage. To reset calibration tests, use the support
software File Control
(p. 539)
menu to delete .cal files, and then send the
demonstration program again to the CR3000.
Case
: A 1000 Ω strain gage is placed into a resistive bridge at position R1. The
resulting circuit is a quarter-bridge strain gage with alternate shunt-resistor (Rc)
positions shown. Gage specifications indicate that the gage factor is 2.0 and that
with a 249 kΩ shunt, measurement should be about 2000 microstrain.
Send CRBasic example FieldCalStrain() Calibration
(p. 245)
as a program to a
CR3000 datalogger.
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......