Section 7. Installation
199
For the example above, the delay is:
Delay in time = (1 ms) • (4
–
1) / 2 = 1.5 ms
Example:
An accelerometer was tested while mounted on a beam. The test had the
following characteristics:
o
Accelerometer resonant frequency ≈ 36 Hz
o
Measurement period = 2 ms
o
Running
average duration = 20 ms (frequency of 50 Hz)
Normalized resonant frequency was calculated as follows:
36 Hz / 50 Hz = 0.72
SIN(0.72π) / (0.72π) = 0.34.
So, the recorded amplitude was about 1/3 of the input-signal amplitude. A
CRBasic program was written with variables Accel2 and Accel2RA. The
raw measurement was stored in Accel2. Accel2RA held the result of
performing a running average on the Accel2. Both values were stored at a
rate of 500 Hz. Figure Running-Average Signal Attenuation
(p. 200)
shows the
two variables plotted to illustrate the attenuation. The running-average value
has the lower amplitude.
The resultant delay, D
r
, is calculated as follows:
D
r
= (scan rate) • (N–1)/2 = 2 ms (10–1)/2
= 9 ms
D
r
is about 1/3 of the input-signal period.
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......