The three-wire half-bridge compensates for lead-wire resistance by
assuming that the resistance of wire a is the same as the resistance of
wire b (see FIGURE: PT100 BrHalf3W() Three-Wire Half-Bridge
Schematic
(p. 281))
. The maximum difference expected in wire resistance
is 2%, but is more likely to be on the order of 1%.
•
The average resistance of 22 AWG wire is 16.5 Ω per 1000 feet, which
would give 500 ft lea
d wires (for example) a nominal resistance of 8.3 Ω.
Two percent of 8.3 Ω is 0.17 Ω. Assuming that the greater resistance is
in wire b, the resistance measured for the PRT in the ice bath (RS0) is
100.17 Ω, and the resistance at 40 °C (RS) is 115.71 Ω.
•
At 40 °C, because of the error from wire b, the measured ratio RS/RS0 is
1.1551 while the ratio without the error would be 115.54/100 = 1.1554.
As a result, the temperature computed by PRTCalc() from the ratio with
the error is about 0.43 °C higher than the temperature measured without
the error from wire b. This source of error does not exist in a four-wire
half-bridge configuration.
7.7.17.5
Example: 100 Ω PRT in
Four-Wire Full Bridge with Voltage
Excitation (PT100 / BrFull() )
FIGURE 69: PT100 BrFull() Four-Wire Full-Bridge Schematic
Procedure
1. Build circuit
1
:
a. Use FIGURE: PT100 BrFull() Four-Wire Full-Bridge Schematic
(p. 285)
as a
template.
b. Choose a 1%, 10 ppm/°C, 5000000 Ω (5 kΩ) resistors for R1 and R4
c. Balance the bridge.
i. Find the midpoint of the temperature range. The range of –40 to 60
°C is selected for this procedure, so the midpoint is 10 °C.
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......