![Campbell CR3000 Micrologger Скачать руководство пользователя страница 233](http://html1.mh-extra.com/html/campbell/cr3000-micrologger/cr3000-micrologger_operators-manual_489090233.webp)
Section 7. Installation
233
•
Two-point slope only
•
Zero basis (designed for use with static vibrating wire measurements)
These demonstration programs are provided as an aid in becoming familiar with
the FieldCal() features at a test bench without actual sensors. For the purpose of
the demonstration, sensor signals are simulated by CR3000 terminals configured
for excitation. To reset tests, use the support software File Control
(p. 539)
menu
commands to delete .cal files, and then send the demonstration program again to
the CR3000. Term equivalents are as follows:
"offset" = "y
-
intercept" = "zero"
"multiplier" = "slope" = "gain"
7.7.12.5.1
FieldCal() Zero or Tare (Opt 0) Example
Most CRBasic measurement instructions have a multiplier and offset parameter.
FieldCal() Option 0 adjusts the offset argument such that the output of the sensor
being calibrated is set to the value of the FieldCal() KnownVar parameter, which
is set to 0. Subsequent measurements have the same offset subtracted. Option 0
does not affect the multiplier argument.
Example Case: A sensor measures the relative humidity (RH) of air. Multiplier
is known to be stable, but sensor offset drifts and requires regular zeroing in a
desiccated chamber. The following procedure zeros the RH sensor to obtain the
calibration report shown. To step through the example, use the CR1000KD
Keyboard/Display or software numeric monitor
(p. 547)
to change variable values as
directed.
Calibration Report for Relative Humidity Sensor
CRBasic Variable
At Deployment
At 30-Day Service
SimulatedRHSignal
output
100 mV
105 mV
KnownRH (desiccated
chamber)
0 %
0 %
RHMultiplier
0.05 % / mV
0.05 % / mV
RHOffset
-5 %
-5.25 %
RH
0 %
0 %
1. Send CRBasic example FieldCal() Zero
(p. 234)
to the CR3000. A terminal
configured for excitation has been programmed to simulate a sensor output.
2. To place the simulated RH sensor in a simulated-calibration condition (in the
field it would be placed in a desiccated chamber), place a jumper wire between
terminals VX1 and SE1. The following variables are preset by the program:
SimulatedRHSignal
=
100, KnownRH
=
0.
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......