Section 7. Installation
286
ii. Select a 1% resistor for R2 with a resistance that is approximately
equal to the resistance of the PRT at 10 °C. See Procedure Information
(PT100 BrFull() Full Bridge)
(p. 296).
Since a 103.9 Ω resistor is hard to
find, use a 100 Ω resistor. It is close enough. Use 5 ppm/°C resistors.
Frequently, all the resistors in a full bridge are submerged in the medium
to be measured, so they may see large temperature changes. 5 ppm
resistors are more thermally stable than 10 ppm resistors.
2. Wire circuit to datalogger:
Use FIGURE: PT100 BrFull() Four-Wire Full Bridge Schematic
(p. 285)
as the
wiring diagram.
3. Calculate excitation voltage:
Use the following equation to calculate the best excitation voltage (VX) for
the measurement range –40 to 60 °C. The equation reduces the absolute
result by 1% to allow for resistor inaccuracy:
VX
max
= (VS
max
/ ((RS
max
/ (RS
max
+ R4)) – (R2 / R1 + R2))) • 0.99
where,
VS
max
= 50 mV (maximum voltage in the
±
50
input range)
R1 = 5000000 mΩ (5 kΩ)
R2 = 100000 mΩ (100 Ω)
R4 = 5000000 mΩ (5 kΩ)
RS
max
= 123240 mΩ (PT100 at 60 °C)
2
so,
VX
max
= 44972562111243 mV
4. Calibrate the PT100:
If the PRT accuracy specification is good enough, and you trust it, assume
RS
0
= 100000 mΩ. Otherwise, do the following procedure:
a. Enter
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......