![Campbell CR3000 Micrologger Скачать руководство пользователя страница 259](http://html1.mh-extra.com/html/campbell/cr3000-micrologger/cr3000-micrologger_operators-manual_489090259.webp)
Section 7. Installation
259
100000 indicates the sensor model.
1.2 is the sensor version.
101 is the sensor serial number.
SDI-12 Start Measurement Commands
Measurement commands elicite responses in the form:
atttnn
where:
a is the sensor address
ttt is the time (s) until measurement data are available
nn is the number of values to be returned when one or more subsequent D!
commands are issued.
aMv!
Starts a standard measurement. Qualifier v is a variable between 1 and 9. If
supported by the sensor manufacturer, v requests variant data. Variants may
include alternate units (e.g., °C or °F), additional values (e.g., level and
temperature), or a diagnostic of the sensor internal battery.
Syntax:
aMv!
As an example, the response from the command 5M! is:
500410
where:
5 reports the sensor SDI-12 address.
004 indicates the data will be available in 4 seconds.
10 indicates that 10 values will be available.
The command 5M7! elicites a similar response, but the appendage 7 instructs the
sensor to return the voltage of the internal battery.
aC!
Start concurrent measurement. The CR3000 requests a measurement, continues
program execution, and picks up the requested data on the next pass through the
program. A measurement request is then sent again so data are ready on the next
scan. The datalogger scan rate should be set such that the resulting skew between
time of measurement and time of data collection does not compromise data
integrity. This command is new with v. 1.2 of the SDI-12 specification.
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......