Section 8. Operation
378
Calibrate() Instruction Results
Cal()
Array
Field
Descriptions of Array Elements
Typical Value
Differential (Diff)
Single-Ended (SE)
Offset or Gain
±mV Input
Range
Integration
31
SE
Offset
5000
50-Hz Rejection
±25 LSB
32
Diff
Offset
5000
50-Hz Rejection
±25 LSB
33
Gain
5000
50-Hz Rejection
–0.167 mV/LSB
34
SE
Offset
1000
50-Hz Rejection
±25 LSB
35
Diff
Offset
1000
50-Hz Rejection
±25 LSB
36
Gain
1000
50-Hz Rejection
–0.0333 mV/LSB
37
SE
Offset
200
50-Hz Rejection
±30 LSB
38
Diff
Offset
200
50-Hz Rejection
±30 LSB
39
Gain
200
50-Hz Rejection
–0.00667 mV/LSB
40
SE
Offset
50
50-Hz Rejection
±30 LSB
41
Diff
Offset
50
50-Hz Rejection
±30 LSB
42
Gain
50
50-Hz Rejection
–0.00167 mV/LSB
43
SE
Offset
20
50-Hz Rejection
±35 LSB
44
Diff
Offset
20
50-Hz Rejection
±35 LSB
45
Gain
20
50-Hz Rejection
–0.00067 mV/LSB
8.1.2.5 Strain Measurements — Details
Related Topics:
• Strain Measurements — Overview
(p. 74)
• Strain Measurements — Details
(p. 378)
• FieldCalStrain() Examples
(p. 243)
A principal use of the four-wire full bridge is the measurement of strain gages in
structural stress analysis. StrainCalc() calculates microstrain (
µ
ε) from the
formula for the particular strain bridge configuration used. All strain gages
supported by StrainCalc() use the full-bridge schematic. In strain-gage parlance,
'quarter-bridge', 'half-bridge' and 'full-bridge' refer to the number of active
elements in the full-bridge schematic. In other words, a quarter-bridge strain
gage has one active element, a half-bridge has two, and a full-bridge has four.
StrainCalc() requires a bridge-configuration code. The table StrainCalc()
Instruction Equations
(p. 379)
shows the equation used by each configuration code.
Each code can be preceded by a dash (-). Use a code without the dash when the
bridge is configured so the output decreases with increasing strain. Use a dashed
code when the bridge is configured so the output increases with increasing strain.
A dashed code sets the polarity of V
r
to negative.
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......