![Campbell CR3000 Micrologger Скачать руководство пользователя страница 299](http://html1.mh-extra.com/html/campbell/cr3000-micrologger/cr3000-micrologger_operators-manual_489090299.webp)
Section 7. Installation
299
CRBasic Programs and Notes
PT100 Resistance() Four-Wire Full-Bridge Calibration
'This program example demonstrates the calibration of a 100-ohm PRT (PT100) in a four-wire
'full bridge with current excitation. See previous procedure and schematic
'
'Declare constants and variables:
Const
R1 = 5000000
'Value of R1 bridge resistor
Const
R2 = 120000
'Value of R2 bridge resistor
Const
R4 = 5000000
'Value of R4 bridge resistor
Public
X
'Raw output from bridge
Public
RS0
'Calculated PT100 resistance at 0 °C
BeginProg
Scan
(1,Sec,0,0)
'Measure X
'Resistance(Dest,Reps,Range,InputChan,IExChan,MPE,ExuA,RevEx,RevDiff,
' SetttlingTime,
Integ
,Mult,Offset)
Resistance
(X,1,mV50,1,Ix1,1,2,True,True,0,250,1,0)
'Calculate RS0:
RS0 = ((-R2*R4) - (X*(R1+R2+R4))) / (X-R1)
NextScan
EndProg
PT100 Resistance() Four-Wire Full-Bridge Measurement
'This program example demonstrates the measurement of a 100-ohm PRT (PT100) in a four-wire
'full bridge with current excitation. See previous procedure and schematic.
'
'Declare constants and variables:
Const
R1 = 5000000
'Value of R1 bridge resistor
Const
R2 = 120000
'Value of R2 bridge resistor
Const
R4 = 5000000
'Value of R4 bridge resistor
Const
RS0 = 100000
'Resistance of PT100 at 0 °C from calibration program
Public
X
'Raw output from bridge
Public
RS
'Calculated PT100 resistance
Public
RS_RS0
'Calculate ratio of RS/RS0
Public
DegC
'Calculated temperature of PT100
BeginProg
Scan
(1,Sec,0,0)
'Measure X
'Resistance(Dest,Reps,Range,InputChan,IExChan,MeasPEx,ExuA,RevEx,RevDiff,
' SettlingTime,
Integ
,Mult,Offset)
Resistance
(X,1,mV50,1,Ix1,1,2,True,True,0,250,1,0)
'Calculate RS and RS_RS0:
RS = ((-R2*R4) - (X*(R1+R2+R4))) / (X-R1)
RS_RS0 = RS/RS0
....'Calculate temperature from RS_RS0:
'PRTCalc(Dest,Reps,Source,PRTType,Mult,Offset)
PRTCalc
(DegC,1,RS_RS0,1,1.0,0)
NextScan
EndProg
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......