Section 8. Operation
382
Analog Voltage Input Ranges and Options
Range Code
Description
2
Append with C to enable common-mode null / open-input detect (Example:
mV200C)
Input Limits / Common-Mode Range
Related Topics:
• Voltage Measurements — Specifications
• Voltage Measurements — Overview
(p. 68)
• Voltage Measurements — Details
(p. 380)
Note This section contains advanced information not required for normal
operation of the CR3000.
Summary
• Voltage input limits for measurement are ±5 Vdc. Input Limits is the
specification listed in Specifications
(p. 95).
• Common-mode range is not a fixed number. It varies with respect to
the magnitude of the input voltage.
• The CR3000 has features that help mitigate some of the effects of
signals that exceed the Input Limits specification or the common-mode
range.
With reference to the figure PGIA with Input-Signal Decomposition
(p. 383),
the
PGIA processes the voltage difference between V+ and V–. It ignores the
common-mode voltage, or voltages that are common to both inputs. The figure
shows the applied input voltage decomposed into a common-mode voltage (V
cm
)
and the differential-mode component (V
dm
) of a voltage signal. V
cm
is the
average of the voltages on the V+ and V– inputs. So, V
cm
= (V+ + V
–)/2 or the
voltage remaining on the inputs when V
dm
= 0. The total voltage on the V+ and
V–
inputs is given as V+ = V
cm
+ V
dm
/2, and V–
=
V
cm
– V
dm
/2, respectively.
The PGIA ignores or rejects common-mode voltages as long as voltages at V+
and V– are within the Input Limits specification, which for the CR6 is ±5 Vdc
relative to ground. Input voltages wherein V+ or V–, or both, are beyond the ±5
Vdc limit may suffer from undetected measurement errors. The Common-Mode
Range defines the range of common-mode voltages that are not expected to
induce undetected measurement errors. Common-Mode Range is different than
Input Limits when the differential mode voltage in non-negligible. The following
relationship is derived from the PGIA figure as:
Common-
Mode Range = ±5 Vdc
– |V
dm
/2|.
Содержание CR3000 Micrologger
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ......
Страница 30: ......
Страница 34: ......
Страница 36: ......
Страница 96: ......
Страница 200: ...Section 7 Installation 200 FIGURE 42 Running Average Frequency Response FIGURE 43 Running Average Signal Attenuation ...
Страница 485: ...Section 8 Operation 485 8 11 2 Data Display FIGURE 110 Keyboard and Display Displaying Data ...
Страница 487: ...Section 8 Operation 487 FIGURE 112 CR1000KD Real Time Custom ...
Страница 488: ...Section 8 Operation 488 8 11 2 3 Final Storage Data FIGURE 113 Keyboard and Display Final Storage Data ...
Страница 489: ...Section 8 Operation 489 8 11 3 Run Stop Program FIGURE 114 Keyboard and Display Run Stop Program ...
Страница 491: ...Section 8 Operation 491 FIGURE 116 Keyboard and Display File Edit ...
Страница 495: ...Section 8 Operation 495 Low power standby whenever possible Low power bus sets bus and modules to low power ...
Страница 496: ......
Страница 502: ......
Страница 564: ...Section 11 Glossary 564 FIGURE 126 Relationships of Accuracy Precision and Resolution ...
Страница 566: ......
Страница 594: ......
Страница 598: ......
Страница 600: ......
Страница 602: ......
Страница 624: ......
Страница 642: ......
Страница 643: ......