![ASL INTERCOM CTR5000 Скачать руководство пользователя страница 95](http://html.mh-extra.com/html/asl-intercom/ctr5000/ctr5000_manual_2984396095.webp)
CTR5000 Operating Instructions
ASL operating instructions, model CTR5000
95
8.3.9
H command
Function:
Holds the current reading, or clears the hold command if already set.
Syntax:
H
Parameters:
None
Initial State:
OFF
Note:
Commands requesting a result or channel change will turn hold off
8.3.10
Hn command
Function:
Holds the current reading, or clears the hold command according to the
parameter following the H
Syntax:
H1 or H0
Parameters:
1 or 0 (H1 = ON; H0 = OFF)
Initial State:
OFF
Note:
Commands requesting a result or channel change will turn hold off
8.3.11
I command
Function:
Set the Probe Id string
Syntax:
Ims--s, where m >= 1 and m <= 72 (probe number)
Parameters:
Sets the Probe Id to s--s (10 characters maximum)
Note:
If the first character of the string is numeric, then the probe number must be two
digits long (e.g. for Probe 2, use I02s-s)
8.3.12
L command
Function:
Local lockout - enables or disables the instrument front panel controls.
Syntax:
Ln, where n = 0 or 1
Parameters:
n = 0 enables front panel (lockout OFF)
n = 1 disables front panel (lockout ON)
Initial State:
n = 0, lockout off
Содержание CTR5000
Страница 102: ...CTR5000 Operating Instructions 102 ASL operating instructions model CTR5000 This page is intentionally left blank ...
Страница 105: ...CTR5000 Operating Instructions ASL operating instructions model CTR5000 105 This page is intentionally left blank ...
Страница 107: ...CTR5000 Operating Instructions ASL operating instructions model CTR5000 107 This page is intentionally left blank ...