17.1
DIN 90 1980 (EN 60751)
The DIN90 standard uses defined values for the conversion of resistance to temperature
using the Callendar van Dusen equation. No data entry for the probe is required since the
coefficients are predefined. The DIN coefficients are defined as –
R
0
= 100 ohms
A = 3.90823 x 10
-3
B = -5.7750 x 10
-7
C = -4.1830 x 10
-12
The CTR5000 will provide accurate resistance to temperature conversion based on these
values when DIN90 conversion is selected for a probe.
17.2
DIN 43760 1980 (IEC 751)
The DIN standard uses defined values for the conversion of resistance to temperature using
the Callendar van Dusen equation. No data entry for the probe is required since the
coefficients are predefined. The DIN coefficients are defined as –
R
0
= 100 ohms
A = 3.90802 x 10
-3
B = -5.8020 x 10
-7
C = -4.2735 x 10
-12
The CTR5000 will provide accurate resistance to temperature conversion based on these
values when DIN68 conversion is selected for a probe.