CTR5000 Operating Instructions
26
ASL operating instructions, model CTR5000
3.5.3
2 Wire PRT
Connect pin 1 to pin 2, pin 4 to pin 5, connect the 2-wire PRT between pin 2 and pin 4.
3.5.4
4 Wire PRT
Connect PRT I+ lead to pin 1, V+ lead to pin 2, V- lead to pin 4 and I- lead to pin 5.
3.5.5
4 Wire PRT with Smart Probe
Refer to document ASL-SMP-10-101 Smart Probe Conversion Kit.
3.5.6
Internal 100Ω check resistor
The CTR5000 contains an internal check resistor. Select channel 99 to use this resistor (IntR will
displayed at the top left of the LCD). It is quite accurate, but not as temperature-stable
11
as
the internal reference resistor. The resistor will be deselected when the channel is changed or
on power-on. It is not necessary to remove any external resistors when using the internal
resistor. The resistor is checked at power-on to ensure the CTR5000 is functioning correctly.
Channel 99 only functions with resistance units.
11
10ppm/
C