CTR5000 Operating Instructions
ASL operating instructions, model CTR5000
89
Type 'help' for help or 'q' to quit.
: ibdev
enter board index: 0
enter primary address: 3 <<= Default for CTR5000!!!
enter secondary address: 0
enter timeout: 13
enter 'EOI on last byte' flag: 1
enter end-of-string mode/byte: 10
ud0: ibwrt “Q_”
<<= <CR> required i.e. \x0D
[0900] ( rqs cmpl )
<<= rqs also
count: 11
ud0: ibrd
enter byte count: 6
<<= line added unless you specify the
<<= number of byte top read it requests
[0900] ( rqs cmpl ) <<= Actual flags
count: 6
30 33 2c 30 31 0d 03,01.
<<= <CR> appears as a dot & is 0D hex
<<= plus ascii in lower case
Note
When the IEEE address is changed (via the Comms Options menu), the
instrument must be switched off and on to reset and use the new
address.