CTR5000 Operating Instructions
114
ASL operating instructions, model CTR5000
will be of special importance to metrology standards departments which need to
make precise measurements at ambient temperatures.
The part of the ITS-90 scale which may be measured by PRTs extends from 83.8058
K (-189.3442
C) to 961.78
C. The actual range of temperatures which may be
measured depends on the type and range of the PRT used.
The ITS-90 scale requires the entry of 6 constants into the CTR5000 before the instrument can
accurately convert resistance to temperature for a particular PRT.
The constants are conventionally called
R
0.01
, Ap, Bp, Cp, Dp, An, Bn
and
W.