CTR5000 Operating Instructions
84
ASL operating instructions, model CTR5000
6.
Calibrating the CTR5000
6.1
CTR5000 reference calibration
The CTR5000 bridge measurement technique used is inherently very stable and linear.
However a small drift of the internal reference resistor may occur with time, making periodic
re-calibration advisable.
Check calibration daily before and after use with known, regularly-calibrated, reference
resistors.
6.2
Equipment
Temperature controlled environment at +20°C ±2°C.
Stable, calibrated (1ppm or better) resistors (0Ω, 100 Ω).
6.3
Calibration procedure
See the section on calibration for the full procedure.
6.4
Full instrument calibration procedure
Special equipment is required for this. Refer to the supplier or agent for details.