CTR5000 Operating Instructions
64
ASL operating instructions, model CTR5000
A spread of up to ±1 Ω is allowed for the resistors. Use the editor to set the calibrated value
and then use the
enter
key (
) to accept this value or the
Esc
key to leave the current value
unchanged.
4.6.1.2
Calibration method
First, enter the calibration value for the 100 Ω resistor from its calibration certificate (see
previous section).
Menu item 2 has been selected (100 Ω calibration) to illustrate the procedure –
Note
Allow at least 40 minutes with power applied before
recalibration.
Follow the on-screen instructions, noting that the relevant external resistor must be connected
to channel A.
Use the
enter
key (
) to accept this step and move on, or press the
Esc
key twice to abort
the procedure
28
. Using the
enter
key, the screen will change to –
28
Use the Esc soft-key to abort the procedure at any time.
1
2
3
4
5
6
7
8
9
0
●
Add 100R ref to ChA
100R Calibration
●
Press enter when ready
Flashing LED