CTR5000 Operating Instructions
66
ASL operating instructions, model CTR5000
Use the
enter
key (
) to accept this value, or press the
Esc
key ignore the calibration
measurement. The last line of the display will briefly show “
Ref calibrated OK
”.
Complete all calibration values.
4.6.1.3
Automatic instrument calibration
A CTS5000 switchbox is required with a set of calibrated resistors. Once started, the calibration
sequence is fully automated. Instrument calibration should be carried out yearly by the
supplier or agent. This menu item is password protected.
1
2
3
4
5
6
7
8
9
0
●
Add 100R ref to ChA
100R Calibration
●
Press enter when ready
●
Measuring . . . . . . . . .
●
Averaging value . . . . . . . . . . .
●
Value checked OK
●
Enter key to accept
Flashing LED (fast)