CTR5000 Operating Instructions
86
ASL operating instructions, model CTR5000
7.
Obtaining the best from the CTR5000 bridge
7.1
Procedure
Please observe the following to obtain the best results from your instrument –
Do ensure there are no draughts or forced air flow around the instrument (e.g. from an
air conditioning unit).
Allow at least 20 minutes for the instrument to fully stabilise, or longer if switching on
from cold.
Do keep the operating environment stable.
Do ensure the sensor leads and input connectors are free from dirt as leakage current
will affect the results.
In common with best practice, use a known stable, calibrated resistor to take a
measurement at the start and end of the day; log the results.
Do NOT locate the instrument near a source of vibration, heat, or draughts.
Do NOT locate the instrument near sources of electrical interference – e.g. electric
motors.