CTR5000 Operating Instructions
ASL operating instructions, model CTR5000
103
10.
Specifications
10.1
Resistance thermometer measurement
39
PRT characterization
ITS90, Din90 BS EN60751:1996, IEC60751:1983
CvD BSEN1904:1984, IEC751:1983
Resistance measurement range
0 to 420
Temperature measurement range
ITS90 -201 to +962ºC
Din90 -201 to +858ºC
CvD -150 to +850ºC
Measurement interval
0.5 seconds
Display resolution
1 mK (0.1 m
when in resistance mode)
Accuracy (resistance)
± 4m
(+20°C ±5°C)
Accuracy (temperature)
± 10mK (+20°C ±5°C)
User selectable measurement display
ºC, ºF, K or ohms
PRT current
1 mA constant current
Internal reference resistor
100
Internal reference resistors thermal stability
0.1ppm/
C
Internal reference resistors accuracy
0.01% (uncalibrated)
Input channels
2, (4 or 6 optionally)
Input connection
5 pin DIN
Expansion port (optional)
CTS5000-8, CTS5000-16 (4 maximum)
39
Measurements do not include uncertainties to the sensor. A warm-up time of 20 minutes is required to meet specification