20
U1401B User’s and Service Guide
1
Getting Started
Display selection with the Hz key
The frequency measurement function is able to detect the
presence of harmonic currents in neutral conductors and
determines, whether these neutral currents are the result of
unbalanced phases or non- linear loads. Press
to enter
the frequency measurement mode for current or voltage
measurements. The voltage or current values will be
displayed on the secondary display and the frequency values
on the primary display. Press this key again to step through
frequency (Hz), duty cycle (%), or pulse width (ms). This
allows simultaneous monitoring of real- time voltage or
current with frequency, duty cycle, or pulse width.
After you press and hold
for more than one second,
the primary display will revert to voltage or current
measurement values.
Table 1-9
Measurement functions and corresponding display selection with the Hz key
Measurement function
Primary display
Secondary display
AC voltage
Frequency (Hz)
AC voltage
Duty cycle (%)
Pulse width (ms)
DC voltage
Frequency (Hz)
DC voltage
Duty cycle (%)
Pulse width (ms)
AC+DC voltage
Frequency (Hz)
AC+DC voltage
Duty cycle (%)
Pulse width (ms)
AC current
Frequency (Hz)
AC current
Duty cycle (%)
Pulse width (ms)
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...