Calibrator Output Operations
Enabling and Disabling the Output
The U1401B can generate and measure signals
simultaneously. Pressing the OUTPUT key disables the
U1401B output by placing it in the standby mode. Pressing
this key again toggles the output on.
When the output is in the standby mode, the
annunciator disappears and the
annunciator is
displayed instead. This means the calibrator has stopped
generating its output.
The standby mode will also be activated automatically if:
•
you accidentally feed an external signal into the output
terminals while the output function is enabled.
•
the noise from an external power system or output
terminals causes an error signal to the output. For
example, when ESD is performed with a voltage of
8000 V, this instrument will go into standby mode.
•
an overload condition has been detected when generating
constant voltage or square wave outputs.
•
the weak or low battery condition occurs. This ensures
output quality and serves as another alert to let the user
know that the energy level of the batteries is low.
•
you put the slide switch into the
M
(input only) position
(you should do this to conserve battery power if you do
not intend to use any of the output functions).
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...