Specifications
8
U1401B User’s and Service Guide
155
Output Specifications
Accuracy is given as ± (% of output
+
counts of least
significant digit) at 23 °C ± 5 °C, with relative humidity less
than 80% R.H., and warmed up for at least five minutes.
Constant voltage and constant current outputs
[1]
The maximum input voltage protection is 30 VDC.
[2]
Loading coefficient: 0.012 mV/mA for 1.5 V output.
[1]
The maximum input voltage protection is 30 VDC.
[2]
Loading coefficient: 1
μ
A/ V, the minimum output voltage is based on 20 mA into a 600
Ω
load.
[3]
If the current loop has a 24 V power, a minimum output voltage of 24 V is achievable with a 20 mA current in a 1200
Ω
load.
This is only applicable for the simulation mode for mA output, refer to
“Simulation Mode for mA Output”
on page 88 for more
information.
Table 8-16
Constant voltage (CV) output specifications
Function
Range
Resolution
Accuracy
Maximum output
current
[2]
Constant
voltage (CV)
[1]
± 1.500 V
0.1 mV
0.03% + 3
25 mA or below
± 15.000 V
1 mV
Table 8-17
Constant current (CC) output specifications
Function
Range
Resolution
Accuracy
Maximum output
voltage
[2]
Constant
current (CC)
[1]
± 25.000 mA
1
μ
A
0.03% + 5
12 V or below
[3]
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...