Maintenance
6
U1401B User’s and Service Guide
121
Recharging the batteries
This instrument is powered by four sets of rechargeable bat-
teries. Charge the batteries immediately when the low bat-
tery annunciator starts flashing. It is strongly recommended
that you use only the specified type of 24 V AC adapter to
charge these rechargeable batteries. Do not turn the rotary
switch while the instrument is being recharged because a DC
24 V supply is applied to the charging terminal.
Follow the procedures below to recharge the batteries:
1
Power down the instrument and disconnect all the test
leads from the terminals.
2
Plug the AC adapter into the jack on the side panel.
3
Set the slide switch at the
CHARGE
position.
4
The red light indicates that the batteries are being
recharged.
5
When the batteries have been fully recharged, the green
light turns on. Remove the AC adapter and set the slide
switch at the
M
or
M/S
position.
WA R N I N G
Do not discharge a battery by shorting it or subjecting it to reverse polarity.
Do not mix different types of battery. Make sure a battery is rechargeable
before charging it.
N O T E
•
A new rechargeable battery comes in a discharged condition and must
be charged before use. Upon initial use (or after a prolonged storage
period), the rechargeable battery may require three to four
charge/discharge cycles before achieving maximum capacity. To
discharge, simply run the instrument using the rechargeable battery's
power until it shuts down or until the low battery warning appears.
•
In some instances, the instrument may indicate that the charging is
complete after ten minutes or so when charging a new rechargeable
battery. This is a normal phenomenon with rechargeable batteries.
Remove the rechargeable battery from the device, reinsert it, and
repeat the charging procedure.
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...