Performance Tests and Calibration
7
U1401B User’s and Service Guide
139
Adjustment Procedures
Input calibration
1
Set the slide switch to the
M/S
position.
2
Allow the instrument to warm up for 20 minutes before
performing calibration.
3
To enter the calibration mode, press
and
for
more than one second. The primary display will indicate
CHEEP
.
4
Press
to enter the input calibration mode.
5
Turn the rotary switch to the “
Test Function
” position as
shown in the adjustment table.
6
The primary display will show the corresponding
reference value of the Cal item.
7
Configure each Cal item.
8
Use the
or
arrow keys to select the Cal range.
9
Apply the input signal shown in the Input column of the
table.
10
The secondary display will now show the value of the
input voltage that is needed; press the HOLD button to
switch between the Offset and Gain values.
11
Repeat steps 5 through 10 for each adjustment point.
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...