Performance Tests and Calibration
Output calibration
1
Set the slide switch to the
M/S
position.
2
Allow the instrument to warm up for 10 minutes before
performing calibration.
3
To enter the calibration mode, press
and
for
more than one second. The primary display will indicate
“CHEEP”.
4
Press
to enter the input calibration mode.
5
Turn the rotary switch to any one of the “Current
Input/Voltage Output” positions, and press
SHIFT
for more
than one second to enter the output calibration mode.
CAL-0 & CAL-1
In the output calibration mode, the primary and secondary
display will show “CAL- 0” and “- rdy- ” respectively.
Connect the output terminals to a multimeter (refer to
Table 7- 1
on page 131 for the recommended test
equipment).
•
CAL-0
:
1
Press
OUTPUT
. The primary and secondary displays
show “CAL- 0” and “00000” respectively.
2
Wait until the instrument reading becomes stable, then
record the value.
•
CAL-1
:
1
Press
MODE
. The primary and secondary displays show
“CAL- 1” and “- rdy” respectively.
2
Press
OUTPUT
. The primary and secondary displays
show “CAL- 1” and “00000” respectively.
3
Press
or
to adjust the output voltage until the
reading on the meter is the same as the
CAL-0
value
recorded above.
4
Press
MODE
to finish the
CAL-0
and
CAL-1
calibration.
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...