148
U1401B User’s and Service Guide
8
Specifications
[1]
Always use the relative function to offset the thermal effect before measuring the signal. If this function is not used, the
accuracy will be 0.03% + 25. The thermal effect could be present in the following conditions:
•
Constant current, constant voltage, or square wave output.
•
Wrong operation — where the resistance, diode, or mV measurement function is used to measure high voltage signals
exceeding 250 V.
•
After battery charging has completed.
•
After measuring a current greater than 50 mA.
AC specifications
[1]
Input impedance: 1.1 M
Ω
in parallel with <100 pF (nominal) for the range of 5 V and above, and 1 G
Ω
(nominal) for the
50/500 mV range. Crest factor:
≤
3.
Table 8-2
DC current specifications
Function
Range
Resolution
Accuracy
Burden voltage/shunt
Overload protection
DC current
50 mA
[1]
1
μ
A
0.03% + 5
0.06 V (1
Ω
)
250 V, 630 mA
Quick acting fuse
500 mA
[1]
10
μ
A
0.6 V (1
Ω
)
Table 8-3
AC mV/voltage specifications
Function
Range
Resolution
Accuracy
Overload protection
45 Hz to 5 kHz
5 kHz to 20 kHz
AC mV/voltage
[1]
(True-rms: From 5%
to 100% of range)
50 mV
1
μ
V
0.7% + 40
1.5% + 40
250 Vrms
500 mV
10
μ
V
0.7% + 20
1.5% + 20
5 V
0.1 mV
50 V
1 mV
250 V
10 mV
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...