Calibrator Output Operations
2
U1401B User’s and Service Guide
35
Defining the autoscan parameters in the memory
Press and hold
MODE
for more than one second to enter the
autoscan adjustment mode. A total of 16 steps with
individually definable time interval and amplitude are
available.
When the instrument is in the autoscan adjustment mode,
the secondary display shows the amplitude. The first two
digits of the primary display are used to indicate which step
is being adjusted. The last two digits of the primary display
are used to indicate time interval.
1
Press
MODE
to cycle through step, time interval, and
amplitude adjustments. The digit to be adjusted will flash
on the display.
•
For amplitude adjustment, press
and
to select
the digit to be adjusted, and then press
and
to
adjust the value of the selected digit. The amplitude
can be set to any value within the selected output
range (±1.5 V or ±15 V for constant voltage output, ±25
mA for constant current output).
•
For time interval adjustment, press
and
to
select the digit to be adjusted, and then press
and
to adjust the value of the selected digit. The time
interval can be set within the range of 0 to 99 seconds.
•
Press
for more than one second to directly reset
the time interval and amplitude of the present step to
zero.
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...