Calibrator Output Operations
2
U1401B User’s and Service Guide
29
Constant Voltage Operation
The U1401B can generate a constant voltage output in two
different ranges, namely ±1.5 V and ±15 V.
To select the constant voltage output function:
1
Turn the rotary switch to any one of the
(constant
voltage output) positions.
2
Press
SHIFT
to access the shifted operations of the keypad.
The
annunciator will appear on the display.
3
Press
MODE
to cycle through ±1.5 V, ±15 V,
±1.5 V,
±15 V,
±1.5 V, and
±15 V output
modes. Select either ±1.5 V or ±15 V for constant output
(or steady output, as opposed to
autoscan
or
autoramp
outputs, which will be discussed under
“Memory
Generation”
on page 31), depending on the voltage range
you require.
•
Unlike the autoscan and autoramp modes, there is no
special annunciator on the display to indicate constant
voltage (CV) operation.
4
With the instrument in standby mode (you should see the
annunciator on the display; if not, press
OUTPUT
), you
can adjust the amplitude of the output by pressing
and
to select the digit to be adjusted, and then
pressing
and
to adjust the value of the selected
digit.
5
Press
OUTPUT
to start the source output. The
annunciator will appear on the display.
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...