134
U1401B User’s and Service Guide
7
Performance Tests and Calibration
2
(cont)
Turn the rotary switch to
. Press
to select
AC.
Connect the
Normal
Hi-Low
output terminals of
the calibrator to the
U1401B input terminals.
500 mV
500 mVrms
@ 45 Hz
±3.7 mVrms
500 mVrms
@ 5 kHz
±3.7 mVrms
500 mVrms
@ 20 kHz
±7.7 mVrms
5 V
5 Vrms
@ 45 Hz
±37 mVrms
5 Vrms
@ 5 kHz
±37 mVrms
5 Vrms
@ 20 kHz
±77 mVrms
50 V
50 Vrms
@ 45 Hz
±0.37 Vrms
50 Vrms
@ 5 kHz
±0.37 Vrms
50 Vrms
@ 20 kHz
±0.77 Vrms
250 V
250 Vrms
@ 45 Hz
±1.95 Vrms
250 Vrms
@ 5 kHz
±1.95 Vrms
250 Vrms
@ 20 kHz
±3.95 Vrms
Table 7-3
Input performance verification tests (continued)
Step
Test Function
Connection to 5520A
Range
5520A output
Error from
nominal 1
year
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...