146
U1401B User’s and Service Guide
8
Specifications
Measurement Category
The U1401B is intended to be used for measurement under
Measurement Category II, 150 V for altitude up to 2000 m.
Measurement category definitions
Measurement CAT I
Measurements performed on circuits that are not directly
connected to MAINS.
For example, measurements on circuits that are not
derived from MAINS, and specifically protected (internal)
mains- derived circuits.
Measurement CAT II
Measurements performed on circuits that are directly
connected to the low voltage installation.
For example, measurements on household appliances,
portable tools, and similar equipment.
Measurement CAT III
Measurements performed in fixed building installation.
For example, measurements on distribution boards, circuit
breakers, wiring (including cables), bus bars, junction
boxes, switches, socket outlets in fixed installation,
equipment for industrial use, and stationary motors with
permanent connections to fixed installation.
Measurement CAT IV
Measurements performed at the source of the low voltage
installation.
For example, electricity meters, measurements on primary
overcurrent protection devices, and ripple control units.
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...