62
U1401B User’s and Service Guide
3
Making Measurements
Refresh hold (auto trigger)
The refresh hold mode freezes the displayed value until the
reading variation exceeds the specified number of counts.
This function will autotrigger and update the held value with
a new measured value. When a new value is updated, the
instrument will beep once as a notification. The keypad
operation is similar to the operation of data hold mode.
1
Make sure the refresh hold mode is enabled in the setup
mode.
2
Press
to enter the refresh hold mode.
•
The present value will be held and the
DH
annunciator
will appear on the display.
•
It will be ready to hold the new measured value once
the variation of the instantaneous reading exceeds the
preset variation count (defined in setup mode); while it
waits for a stable new reading, the
DH
annunciator will
flash.
•
The
DH
annunciator will stop flashing once a stable
new reading is available, and then the new value will
be updated to the display. The instrument will beep
once as a notification.
3
Press
to exit this mode.
For voltage and current measurements, the held value will
not be updated if the variation of the reading is below 500
counts. For resistance and diode measurements, the held
value will not be updated if the reading is
OL
or open. For
all measurements, the held value will not be updated if the
reading cannot reach a stable state.
Содержание U1401B
Страница 1: ...Agilent Technologies U1401B Handheld Multi Function Calibrator Meter User s and Service Guide ...
Страница 16: ...XVI U1401B User s and Service Guide Contents ...
Страница 39: ...Getting Started 1 U1401B User s and Service Guide 19 The rear panel at a glance Figure 1 8 The rear panel ...
Страница 46: ...26 U1401B User s and Service Guide 1 Getting Started ...
Страница 64: ...44 U1401B User s and Service Guide 2 Calibrator Output Operations ...
Страница 67: ...Making Measurements 3 U1401B User s and Service Guide 47 Figure 3 1 DC voltage measurement ...
Страница 73: ...Making Measurements 3 U1401B User s and Service Guide 53 Figure 3 4 Surface temperature measurement ...
Страница 104: ...84 U1401B User s and Service Guide 4 Changing the Default Settings ...
Страница 115: ...Application Examples 5 U1401B User s and Service Guide 95 Figure 5 5 Zener diode test ...
Страница 134: ...114 U1401B User s and Service Guide 5 Application Examples Figure 5 16 Verifying a two wire transmitter Red Black ...
Страница 136: ...116 U1401B User s and Service Guide 5 Application Examples Figure 5 17 Verifying a frequency transmitter ...
Страница 146: ...126 U1401B User s and Service Guide 6 Maintenance ...